Description: Transducer TSI, VK-P Series are offered by Forbes Marshall from Shinkawa Electric, Japan. The VK-P series TSI transducer is a high-grade, eddy-current type, non-contacting transducer for bearing position and differential expansion measurement required for TSI. It offers the necessary low temperature drift and stability for displacement measurement & can also be used in an explosive gas atmosphere due to its intrinsically safe construction. System composition: consists of sensor, extension cable and driver. The transducer is available in lengths of 5m and 9m, the total distance between sensor & driver. Various type of measured ranges: The VK-P series transducer is offered in five measured ranges of 2, 3, 6, 13.5 and 26mm. It is suitable for long-range measurement as required for TSI. Intrinsic safety explosion-proof construction: TIIS equivalent to IEC standard Ex is IIC T5, CSA, FM, ATEX. Anti-corrosion: extensive operating temperature range (-40 to +125°C). Not affected by lubricating oil or dust (excluding connector). Ammonia resistant and hydrogen sulfide models are also available.
Other Catalogs (From Forbes Marshall Pvt. Ltd.) :